The SEMTECH® PM2 is our 2nd generation on-board Particulate Mass measurement device that combines real-time and cumulative PM measurements. The real-time PM mass measurement is based on an ion mobility technique, and the cumulative PM mass measurement is based on the gravimetric method using industry standard 47 mm filter cartridges.

Download a Brochure
Request a Quote

SEMTECH PM2

SEMTECH PM2: Particulate Matter Measurement

The system is compatible with both the SEMTECH® LDV and the SEMTECH® ECOSTAR Plus. The PM2 analyzer utilizes an ejector dilution system. HEPA filtered dilution air flows into the diluter and is directed around the ejector nozzle at a high velocity. This causes a pressure drop which draws in the sample through the nozzle. The raw sample mixes with the dilution air resulting in a stable homogeneous diluted sample.

For real-time PM mass measurement, a small fraction of the homogenous diluted sample is passed through a Pegasor PPS-M electrical aerosol transducer, in which particulate matter is charged. As the charged particles exit the detection zone, their charge is measured with a sensitive electrometer. The electrometer signal is then converted to real-time PM mass concentration using a transfer function.

For cumulative PM mass measurement, the homogeneous diluted sample is directed through one of three filters. The pre- and post- filter weights provide cumulative PM mass, which is used to scale the real-time mass concentration and obtain accurate, time resolved PM mass measurement

  • Features
  • Specifications
  • US EPA 40 CFR Part 1065, Heavy Duty In Use, Alternative Method
  • Ejector dilution system
  • Traceable reference pressure and flow sensors
  • Pegasor PPS-M electrical aerosol transducer
  • Three 47 mm filter cartridges
Measurement Technique Ion mobility (Pegasor PPS-M transducer)
Minimum Particle Size 23 nm (dependent on selected trap voltage)
Maximum Particle Size 2.5 μm (dependent on measured particle size distribution)
Concentration From 1 μg/m3 to 250 mg/m3
Response Time < 0.3 seconds (Dispersed to t10-t90 ca 2.5 seconds)
Maximum Data Rate 5 Hz (high speed internal averaging)
Power Requirement 12 VDC
Power Consumption < 250 W (excluding vacuum and compressor pump)
Storage Temperature Dry -10°C to 60°C
Operating Temperature -10 to 60°C ambient
Operating Environment -10 to 45 °C ambient
0 to 90% RH
Dimensions (WxDxH) 43.6 x 31.1 x 17.7 cm
17.2” x 12.3” x 7.0"
Weight: 20kg
44lbs
Data Transmission Ethernet
Electromagnetic
Interference & Susceptibility
CE Standards: ICE 61326:2002-2