19" Rack FID: 1230
The FID 1230 module measures total hydrocarbons in a wide range of applications, such as catalytic and thermal oxidiser plants, waste gas processing plants, room and environmental air, solvent recovery plants and vehicle exhaust gases. The monitoring is continuous with high accuracy.
- Features
- Specifications
- Options
- Built-in, titanium, heated sample gas filter
- Heated to 200°C, or 300°C option
- Filter monitoring
- Automatic flame ignition
- Hydrogen cut-off
| Measuring component: | CxHy |
| Display: | 6 digit LED |
| Decadic measuring range: | 5 |
| Smallest measuring range: | 0 - 10 ppm |
| Largest measuring range: | 0 - 100.000 ppm |
| Range selection: | manual / automatic |
| Repeatability: | ±1% of reading |
| Instrument zero drift: | ±1% in 24 hr |
| Response time: | T90 < 1.0 second |
| Warm-up time: | approx. 30 minutes |
| Analog output current loop: | 0 - 20 mA or 4 - 20 mA |
| Analog output voltage: | 0 - 10 V |
| Power supply: | 115V / 60 - 50 Hz |
| Power consumption: | 300 W |
| Ambient temperature: | 0 - 45°C |
| Size (H x D): | 3U x 460mm |
| Weight: | 23 kg |
Gas Requirements |
|
| Fuel: | H2, 5.0 |
| Fuel consumption: | approx. 35 ml/min |
| Span gas: | C3H8 |
| Span gas consumption: | 1 l/min |
| Zero gas: | N2, 5.0 or synth. air |
| Zero gas consumption: | 1 l/min |
| Combustion air: | built in catalyst |
- USB or RS232 interface
- Temperature controlling module
- Control module for temperature, pressure and H2
- Switch module for automatic measurement range
- FID 1230P with membrane pump
- FID 1230I with air jet pump (compressed air required)
- Table housing