The FID 2000MP analyzes hydrocarbons in SHED chambers of varying dimensions. For applications over 2.5 m3, we recommend an analyzer with a heated bypass system to ensure accurate response times. For smaller SHED chambers, we offer an analyzer that minimizes the air change rate in the chamber.

 

  • Features
  • Specifications
  • Options
  • Resources
  • Optimal sample extraction
  • 4 span gas inlets
  • Automatic calibration
  • Automatic flame ignition
  • Digital output

Measuring component: CxHy
Measuring ranges: 4
Smallest measuring range: 0 - 10 ppm
Largest measuring range: 0 - 100.000 ppm
Range selection: manual / remote
Repeatability: ±1% of reading
Instrument zero drift: ±1% in 24 hr
Response time: T90 < 1.0 second
Warm-up time: 20°C to 191°C approx. 30 minutes
Analog output current:: 0 - 20 mA or 4 - 20 mA
Analog output voltage: 0-10V
Alarm: Flame control
Vacuum FID: 0.4 bar vacuum
Power supply: 230V / 50 Hz
Power capacity: 600 W
Ambient temperature: 0 - 45°C
Size (H x D): 3U x 460mm
Weight: 23 kg
Gas Requirements
Fuel: He/H2
Fuel consumption: 150 ml/min
Span gas: C3H8
Span gas consumption: 500 ml/min
Zero gas: N2, 5.0
Zero gas consumption: 500 ml/min
Combustion air: from room air, over active coal
Combustion air consumption: 1000 ml/min
  • Heated lines
  • Zero gas generator
  • Unheated filter for inside SHED chamber
  • Heated solenoid valve