The FID 2010 T measures total hydrocarbons in a wide range of applications, such as catalytic and thermal oxidiser plants, waste gas processing plants, room and environmental air, solvent recovery plants and vehicle exhaust gases. A compact and lightweight design, it is ideal for short term measurements, and measuring points that frequently change.
- Features
- Specifications
- Options
- Resources
- Built-in, titanium, heated sample gas filter
- Heated to 200°C, or 300°C option
- Filter monitoring
- Automatic flame ignition
- Hydrogen cut-off
| Measuring component: | CxHy |
| Display: | 6 digit LED |
| Decadic measuring range: | 5 |
| Smallest measuring range: | 0 - 10 ppm |
| Largest measuring range: | 0 - 100.000 ppm |
| Range selection: | manual |
| Repeatability: | ±1% of reading |
| Instrument zero drift: | ±1% in 24 hr |
| Response time: | T90 < 1.0 second |
| Warm-up time: | approx. 25 minutes |
| Analog output current loop:: | 0 - 20 mA or 4 - 20 mA |
| Analog output voltage: | 0 - 10 V |
| Power supply: | 115V / 60 - 50 Hz |
| Power consumption: | 300 W |
| Ambient temperature: | 0 - 45°C |
| Size (H x W x D): | 200mm x 410mm x 420mm |
| Weight: | 15 kg |
Gas Requirements |
|
| Fuel: | H2, 5.0 |
| Fuel consumption: | approx. 35 ml/min |
| Span gas: | C3H8 |
| Span gas consumption: | 1 l/min |
| Zero gas: | N2, 5.0 or synth. air |
| Zero gas consumption: | 1 l/min |
| Combustion air: | catalysator built-in |
- Carrying case
- USB or RS232 interface
- Temperature controlling module
- Control module for temperature, pressure and H2
- Switch module for automatic measurement range