The Advanced Gasmitter (AGM) series uses the latest technologies in both Non-Dispersive Infra-Red (NDIR) and thermally conductive gas transducers for industrial applications. Both types offer high accuracy, long term stability and excellent repeatability. The flexible design platform enables easy integration for the system developer.
- AGM Products
- Features
- Specifications
- Options
- Resources
AGM Series
Reliable Gas Measurement
- AGM 10: CO2 / CH4
- AGM 30: CO2 / CH4
- AGM 31: CO2 / N2O
- AGM 32: CO2 / N2O / C3H8
- AGM 50: CO2
- AGM 101: CO2 / CH4 / C6H14 / N2O in low concentrations
- AGM 106: CO
- AGM 22: H2 / He / Xe / Ne / Ar / SF6 (uses thermal conductivity in lieu of NDIR)
- NDIR-Dual wavelength or thermal conductivity sensor for gas measurement
- Proven and tested for over 13 years
- Linearized analog output (current or voltage)
- Analog output signal with selectable time constant
- Output signal for 2 alarm thresholds
- RS232 Interface
- Zero and endpoint adjustable with potentiometers
- Temperature compensated
- European PCB format (160mm x 100mm)
| General | |
| Measuring principle: AGM 10/30/31/32/50/101/106 |
Non Dispersive Infra-Red (NDIR) dual wavelength |
| Measuring principle: AGM 22 |
Thermal conductivity principle |
| Measurement Response1 | |
| Linearity error: | < ±2% FS* |
| Lower detection limit: | < 1% FS* |
| Repeatability at zero: | ±0.2% FS* |
| Repeatability at span: | ±1% FS* |
| Long term stability: | ±2% FS* (12 months) |
| Influencing Variables1 | |
| Temperature dependence at zero: | ±1% FS* per 10K |
| Temperature dependence at span: | ±2% FS* per 10K |
| Heating option: | 55°C ± 5°C |
| Pressure dependence at zero: | 0% |
| Pressure dependence at span: | < 2% FS* per 10mbar non compensated; <0.2% FS* per 10mbar with optional pressure compensation (in the range of 800 - 1200 mbar) |
| Time Response | |
| Warm up: | 10s operation (≤ 30 min. to approach all specified properties) |
| Response time (T90): | 5 - 45s (depending on used sensor element, flow, tube length, digital attenuation) AGM22 = 1-45s |
| Gas Inlet Conditions | |
| Ambient temperature: | 0 - 50°C in process -20 - 60°C in storage |
| Gas pressure: | 800 - 1200 mbar |
| Ambient humidity: | < 95% RH (non condensing) |
| Flow range (gas): | 0.1 - 0.750 l/min (constant) |
| Calibration | |
| Zero point calibration: | with inert gas (e.g. nitrogen N2, or ambient air free from measurement gases) |
| Span point calibration: | with sample gas mixtures |
| Adjustment: | zero point and span point adjustable with potentiometer on the PCB. |
| Recalibration: | At first initiation and depending on application and ambient conditions. Recurring cycles of recalibration are recommended. |
| Electric Properties | |
| Operating voltage: | 24 VDC (18-30VDC); 24 VDC (24-30VDC) with heating option |
| Power consumption: | Max. 3W; 8W with heating option; 28W during warm up |
| Status / Failure visualization: | Status and failure LED on the PCB |
| Status / Failure output: | +5V HCMOS on 34-pin connector |
| Analog output (linearized): | 0 - 20 mA, 4 - 20 mA;optional voltage output 0 - 1V, 0 - 5V, 0-10V Max. 500Ω load at current output; ≤ 300Ω burden at voltage output |
| Digital output (linearized): | RS232 Interface (V.24/V.28) 12 bit resolution |
| Dimensions / Materials | |
| PCB dimensions: | 160mm x 100mm (European PCB format); Height: 45mm (includes connectors and tubes) |
| Weight: | max. 800g |
| Tube connector: | 3mm (outside diameter); Hose inside diameter < 3mm |
| Materials in contact with medium AGM 10/30/31/32/50/101/106: | Sample cell: Aluminum or gold-plated stainless steel Window: CaF2, Al2O3 or Ge Gas hoses: Viton®(FPM/FKM) Hose connectoirs: nickel-plated brass Filter: Polypropylene (PP) |
| Materials in contact with medium AGM22: | Sample cell: Al, Si, SiOxNy, Au, Epoxy Gas hoses: Viton®(FPM/FKM) Hose connectoirs: nickel-plated brass Filter: Polypropylene (PP) |
| Options | |
Pressure compensation
|
|
Heating option |
|
Certificate of compliance |
|
1 = Refers to 1013 mbar sample gas pressure, 25°C ambient temperature and 0.75 l/min constant gas flow
- Pressure compensation
- Heating option
- Versions for gas diffusion operation available
- Certificate of compliance
- AGM NDIR Brochure (English)
- AGM NDIR Brochure (German)
- AGM Thermal Conductivity Brochure (English)
- AGM Thermal Conductivity Brochure (German)
- AGM Specifications (English & German)
- Request Quote